Insulation resistance measurement:
– measurement voltage any in the range of 50…1000 V at 10 V and 1000…5000 V at 25 V resolution,
– continuous indication of measured insulation resistance or leakage current,
– automatic discharge of measured object capacitive voltage after the end of insulation resistance measurement,
– acoustic signaling of 5 seconds intervals to facilitate capturing time characteristics,
– metered T1 , T2 and T3 test times for measuring one or two absorption coefficients from the range of 1…600 s,
– adjustable measuring time to 99’59’’,
– polarization index (PI) and dielectric absorption ratio (DAR) measurement,
– indication of actual test voltage during measurement,
– 1.2 mA and 3 mA test current,
– step voltage insulation resistance measurement (SV),
– Dielectric Discharge calculation (DD),
– protection against measuring live objects,
– measurements with test leads up to 20 m
Digital filters function for measurements in high noise enviroment (10 s, 30 s , 60 s).
Continuity measurement of protective connections and equipotential bonding in accordance with EN 61557-4 with current > 200 mA.
Adjustable limits for measured resistance RISO and RCONT.
Measurement of leakage current during insulation resistance testing.
Measurement of capacitance during the measurement of RISO.
DC and AC voltage measurement in the range of 0…600 V.
990 cells of memory (11880 records) with the capability of wireless data transmission to a PC (with the USB-OR adapter) or through a USB cable.
Power supply from battery packs, low battery warning indicator, built-in fast charger.
Keyboard and display backlit.
The instruments meet the requirements of the EN 61557 standard.
Insulation resistance measurement (two-lead)
Measurement range acc. to IEC 61557-2: 50k…15,0 TΩ (IISOnom = 1,2 mA or 3 mA)
Range
Resolution
Accuracy
0,0…999 kΩ
1 kΩ
±(3% m.v. + 10 digits)
1.00…9,99 MΩ
0,01 MΩ
10.0…99,9 MΩ
0,1 MΩ
100…999 MΩ
1 MΩ
1.00…9,99 GΩ
0,01 GΩ
10.0…99,9 GΩ
0,1 GΩ
100…999 GΩ
1 GΩ
±(3.5% m.v. + 10 digits)
1.00…9,99 TΩ
0,01 TΩ
±(7.5% m.v. + 10 digits)
10.0…15,0 TΩ
0,1 TΩ
±(10% m.v. + 10 digits)
Values of measured resistance depending on measurement voltage
Voltage UISO
Measurement range
250 V
500 GΩ
500 V
1.00 TΩ
1000 V
2.00 TΩ
2500V
5.00 TΩ
5000V
15.0 TΩ
Measurement of leakage current
Range
Resolution
Accuracy
0…ILmax
m, µ, n [A]
Calculated basing on
resistance measurements
· ILmax – maximum current at short circuit of leads,
· resolution and units result from the measurement range of individual insulation resistance.
Step voltage insulation resistance measurement
Target voltage
Measurement
voltage sequence
1 kV
200, 400, 600, 800, 1000 V
2.5 kV
0.5, 1, 1.5, 2, 2.5 kV
5 kV
1, 2, 3, 4, 5 kV
· duration of each “step” adjustable from 30s to 5mins
· measurement result for each voltage step is stored in memory
Continuity measurement of protective connections and equipotential bonding with 200 mA current
Measurement range acc. to EN 61557-4: 0,12…999 Ω
Range
Resolution
Accuracy
0.00…19.99 Ω
0,01 Ω
±(2% m.v. + 3 digits)
120.0…199.9 Ω
0,1 Ω
200…999 Ω
1 Ω
±(4% m.v. + 3 digits)
· Voltage on open terminals: 4…24 V
· Output current at R < 2 Ω: Imin > 200 mA (ISC: 200…250 mA)
· Compensation of test lead resistance
· Current flowing in both directions, mean value of resistance is displayed
Capacity measurement
Range
Resolution
Accuracy
1…999 nF
1 nF
±(5% m.v. + 5 digits)
1.00…49.99 µF
0.01 µF
· Capacity measurement result is displayed after the RISO measurement
DC and AC voltage measurement
Range
Resolution
Accuracy
0.0…29.9 V
0.1 V
±(2% m.v. + 20 digits)
30.0…299.9 V
0.1 V
±(2% m.v. + 6 digits)
300…600 V
1 V
±(2% m.v. + 2 digits)
The acronym “m.v.” stands for a “measured reference value”.
MIC-5010 Insulation Resistance Meter
Description
– measurement voltage any in the range of 50…1000 V at 10 V and 1000…5000 V at 25 V resolution,
– continuous indication of measured insulation resistance or leakage current,
– automatic discharge of measured object capacitive voltage after the end of insulation resistance measurement,
– acoustic signaling of 5 seconds intervals to facilitate capturing time characteristics,
– metered T1 , T2 and T3 test times for measuring one or two absorption coefficients from the range of 1…600 s,
– adjustable measuring time to 99’59’’,
– polarization index (PI) and dielectric absorption ratio (DAR) measurement,
– indication of actual test voltage during measurement,
– 1.2 mA and 3 mA test current,
– step voltage insulation resistance measurement (SV),
– Dielectric Discharge calculation (DD),
– protection against measuring live objects,
– measurements with test leads up to 20 m
The instruments meet the requirements of the EN 61557 standard.
Insulation resistance measurement (two-lead)
Measurement range acc. to IEC 61557-2: 50k…15,0 TΩ (IISOnom = 1,2 mA or 3 mA)
Values of measured resistance depending on measurement voltage
Measurement of leakage current
resistance measurements
· ILmax – maximum current at short circuit of leads,
· resolution and units result from the measurement range of individual insulation resistance.
Step voltage insulation resistance measurement
voltage sequence
· duration of each “step” adjustable from 30s to 5mins
· measurement result for each voltage step is stored in memory
Continuity measurement of protective connections and equipotential bonding with 200 mA current
Measurement range acc. to EN 61557-4: 0,12…999 Ω
· Voltage on open terminals: 4…24 V
· Output current at R < 2 Ω: Imin > 200 mA (ISC: 200…250 mA)
· Compensation of test lead resistance
· Current flowing in both directions, mean value of resistance is displayed
Capacity measurement
· Capacity measurement result is displayed after the RISO measurement
DC and AC voltage measurement
The acronym “m.v.” stands for a “measured reference value”.